娇妻聚会大杂交
下载文件详细资料
  文件名称: 老化试验箱测试方法
  公司名称: 东莞市宏力仪器有限公司
  下载次数: 961
  文件详细说明:
 
老化试验箱测试方法
  1、升温时间的测试
  将温度传感器置于高温老化试验箱工作空间内任意一点,用全功率加热,记录工作室温度从室温升至次达到高工作温度的时间,其结果应不超过120min。
  2、表面温度测试
  在高温老化试验箱工作温度次达到高工作温度并稳定2h后,用温度计测试箱体表面的温度,其高工作温度不超过200℃的试验箱,表面温度应不大于室温加35℃;高工作温度超过200℃,表面温度应按式决定。
  3、 绝缘电阻测试
  应按GB99 8中第6.2条规定方法进行测试,其结果应符合电加热器端子(包括引线)与控制系统开路时,对箱壳应承受电压1500V,交流50H,历时1min的绝缘强度试验,其绝缘应不被击穿。
  4、绝缘强度试验
  应按GB998中第6.3条规定进行试验,试验结果符合电加热器端子(包括引线)与控制系统开路时,对箱壳应承受电压1500V,交流50H,历时1min的绝缘强度试验,其绝缘应不被击穿。
  5、噪声测试
  按ZBN61012中规定方法进行测试,其结果整机噪声不高干70dB (A )
文件下载(右键文件另存为)